For Figure 4 MED-AMB 10 and MED-WHT 10 materials, it may be necessary to sterilize cured, finished parts to be used in certain medical applications. You can sterilize printed parts by making use of an autoclave or using gamma-ray sterilization. Do not use a dishwasher or thermal disinfector. 3D Systems recommends the following autoclave method as a baseline starting point for sterilization:
NOTE: The above protocol is a recommended starting point and may not be appropriate for all cases. Different geometries and use may require adjustment of this protocol to maximize effectiveness. |
If using an autoclave for sterilization, you will notice that your part will change color with the sterilization process. This is normal. Please see the before/after images below for reference.
![]() |
|
Before sterilization | After sterilization |
![]() |
|
Before sterilization | After sterilization |
NOTE: Autoclave sterilization may impact the mechanical properties. Small drops in tensile strength, elongation at break, and impact strength may occur. Such changes are typically less than 20%. Use cases that are sensitive to that should be validated. |